High performance capillary electrophoresis instrument High performance capillary electrophoresis is a new type of liquid phase analysis technology with capillary as separation channel and high voltage
Details

High performance capillary electrophoresis instrument High performance capillary electrophoresis is a new type of liquid phase analysis technology with capillary as separation channel and high voltage direct current electric field as driving force. It is another major progress in analytical science after high performance liquid chromatography. It enables analytical science to move from micro rise level to nano rise level, and makes single cell and even single molecule analysis possible, Therefore, the separation and analysis of biological macromolecules such as proteins, which have puzzled us for a long time, have a new idea. The instrument can be used to analyze organic compounds, inorganic ions, neutral molecules, chiral compounds, proteins and peptides, DNA and nucleic acid fragments.

1、 Performance parameters

(★ is an important indicator)

1、 Dual wavelength detector (imported from the United States)

1. Optical path system: diffraction grating monochromator

2. Light source: imported deuterium tungsten lamp, edge filtering, 370nm

3. ★ Automatic switching system of deuterium lamp and tungsten lamp

4. ★ Dual wavelength detection wavelength range: 190-740nm

5. Bandwidth: Δλ ≤8nm

6. Wavelength accuracy: ± 1nm

7. Sensitivity: 2 × 10-5AU at 254nm, t=1s

8. GLP function: detailed report, lighting time and lighting times

9. Noise: 1.5 × 10-5AU

10. Baseline drift: 4 × 10-4AU/h

11. ★ Detection limit: ≤ 1 × 10-5g/ml

12. Quantitative repeatability: RSD% ≤ 3.0%

13. Qualitative repeatability: RSD% ≤ 1.3%

14. ★ Minimum capillary length: 25 cm

2、 High voltage power supply

1. ★ Output voltage range: 0-35KV continuously adjustable (standard configuration)

2. Output current range: less than 0-1000 μ A

3. Voltage output error: ≤ 0.5%

4. Current output error: ≤ 0.5%

5. Time drift: ≤ 0.05%

6. Temperature drift: ≤ 0.05%

7. ★ One click high voltage and data acquisition

8. Short circuit protection: closing section self recovery protection

9. Variable wavelength selection, 190-740nm continuously adjustable

10. Injection mode: electric injection 0-35KV adjustable,

11. LCD screen display, touch the panel

12. Wavelength scanning and automatic zeroing